Connections
2 Connections
Caution
Unless appropriate accessories and attach-
ments are used, the TURBOVAC is not
suitable for aggressive or corrosive media,
or those containing dust. When handling
corrosive media the C version pump must
be operated with purge gas; when handling
media containing dust, a fine-mesh filter
must be installed.
Observe the information on media compati-
bility, to be found at the beginning of these
operating instructions.
Do not open the packaging until immediately prior to
installation.
Remove the covers and the blank flanges at the turbo-
molecular pump only just before installing, to ensure that
the TURBOVAC is installed under the cleanest possible
conditions.
When moving the heavier pumps, the lower surfaces on
the base flange are suitable for accepting a lifting fork;
see Fig. 3.
Caution
Take care not to damage the plugs and
coolant connections during movement.
Do not stand below the TURBOVAC pump while it is being
connected to or detached from the system.
8
GA05118_1302 - 01/2004
TURBOVAC
50 D2
50
151, 151 C,
361, 361 C
600 C
1000 C
Air or water cooling required
Ambient temperature 30 to 40 °C
Bakeout operations
High-vacuum pressure 10
-3
to
10
-2
mbar
Forevacuum pressure 5·10
-1
to
2 mbar
Bakeout at ambient temperature
< 40 °C
Continuous operation at high-vacuum
pressure > 10
-4
mbar
Rapid-cycling operation
Bakeout at ambient temperature
< 35 °C
High-vacuum pressure < 10
-3
mbar
and ambient temperature < 45 °C
Bakeout at ambient temperature
< 35 °C or
high-vacuum pressure < 10
-3
mbar
and ambient temperature < 45 °C
Bakeout at ambient temperature
< 35° C or
high-vacuum pressure < 10
-4
mbar
and ambient temperature < 45 °C
Water cooling required if any one of these
conditions prevails
Ambient temperature 40 - 45 °C
Bakeout operations
High-vacuum pressure > 5·10
-2
mbar
Forevacuum pressure > 2 mbar
Ambient temperature 45 - 55 °C
Back-out at ambient temperature > 40° C
Bakeout at ambient temperature > 35 °C
High-vacuum pressure > 10
-3
mbar
Ambient temperature 45 - 55 °C
Bakeout at ambient temperature > 35 °C
High-vacuum pressure > 10
-3
mbar
Ambient temperature 45 - 55 °C
Bakeout at ambient temperature > 35 °C
High-vacuum pressure > 10
-4
mbar
Ambient temperature 45 - 55 °C
No additional cooling required if all
these conditions are satisfied
Ambient temperature < 30° C
No bakeout operations
High-vacuum pressure < 10
-3
mbar
Forevacuum pressure < 5·10
-1
mbar
Continuous operation at high-
vacuum pressure < 10
-4
mbar
Ambient temperature < 45 °C
Operating environment and cooling
Fig. 3 Movement
°C °F
———————
30 86
35 95
40 104
45 113
50 122
55 131
mbar Torr
———————
10
-4
8·10
-5
10
-3
8·10
-4
10
-2
8·10
-3
5·10
-2
4·10
-2
5·10
-1
4·10
-1
2 1.5